f. Press A to scroll to New observation. Then Press
C to measure and record angles
g. Re sight the instrument at the same point and press
the enter key again. Repeat this process for the same
number of times as in face 2.
As observations are made on the first face (either face 1 or
face 2), the angle values are stored and the counter increases.
When one or more observations have been taken on each face,
and the number of observations on each face are the same,
the software calculates and displays the new horizontal and
vertical collimation values.
4. Press A to scroll to one of the following:
– Trunnion coll. Then press C to continue to
Trunnion collimation.
– Store correction. Then press C to accept and
store the new collimation values.
– Cancel. Then press C to return to the adjustments
menu.
Select Trunnion coll. to continue with trunnion axis tilt
collimation.
5. Press A to scroll to Trunnion coll. Then press C to
continue the Trunnion axis tilt test.
Face-1 obs: 3
New observation
Change face
Face-2 obs: 3
>>
HA: 0.0015
VA: 0.0008
-Current values-
Trunnion coll.
>>
Store correction