The Horizontal and Vertical collimation and the trunnion axis
tilt correction have been measured and stored in the
instrument at the factory.
In all calibrations, multiple sightings will be made in both faces
to ensure that any minor pointing errors can be eliminated in
the accurate determination of current collimation error
values.
In a new instrument the values should be close to zero, over
time these will change. The instrument allows a maximum
value of 0.05 grads (0.045 degrees) in the HA, VA and Trunnion
axis tilt values. If these values are exceeded, the instrument
will need service to rectify a mechanical problem.
1. Press A to scroll to HA/VA collimation and then press
C. The current collimation values appear.
2. Press A to scroll to one of the following:
– Continue Then press C to continue the HA/VA
collimation test.
– Cancel. Then press C to return to the Adjustments
menu.
HA: 0.0010
VA: 0.0012
Cancel
-Current values-
Continue
>>
C
-Adjustments-
Tracker collim.
HA/VA collimation
Compensator calib.
Laser pointer
Back...
>>