5. Maintenance
In order to ensure that the equipment is used safely, be sure to inspect the equipment before use. If any
problem is detected during the inspection and cannot be corrected, please contact your sales representative
or local DRTECH dealer.
5.1 Daily Inspection and Maintenance
WARNING
AVERTISS-EMENT
For safety reasons, be sure to turn OFF the power to each piece of equipment
before the following procedures. Otherwise, an electric shock may result.
Cable
- Ensure that cables are not damaged and cable jackets are not torn.
- Ensure that the power cord plugs are securely connected to toth the equipment AC inlet
and the AC outlet.
Detector
- Ensure that there are no loose screws or broken parts.
- Ensure that there is no dust or foreign substance on the external connector.
- Ensure that there are no broken parts or short-circuits in the power supply connector.
After turning on the power
- Be sure to start the ECali1 before performing the inspection.
- Perform exposure testing.
5.2 Regular Inspection and Maintenance
Monthly Inspection
- Ensure that there are no loose screws or broken parts.
- Ensure that there is no dust or foreign substance on the external connector.
Yearly Inspection
- Perform a Performance Test or Self-diagnosis Test using a phantom or resolution chart,
etc.
5.3 Irregular Inspection and Maintenance
Calibration
- Perform Calibration when exposure conditions have changed significantly.
For details, refer to the Setup Guide for ECali1.