User's Manual

DRT-MAN-093
69/74
4. Trouble Shooting
When you encounter problems or error messages in the EVS Console while using this equipment, search the
table below for the problem or error message and try the solutions.
If the problem persists, turn off the detector and consult your sales representative or local DRTECH dealer.
Please refer to the details of the problem or error messages.
Symptom
Cause/Error messages on the
EVS Control Software
Remedy
Failed to turn on.
Not Installing battery pack well
Dead battery pack
Battery pack or Detector is
broken
Install the battery pack.
Charge the battery pack.
Replace other battery packs and
check the results.
Replace other Detectors and check
the results.
Replace corresponding devices.
Ready LED(Orange) lamps
of detector are blinking 3
time/sec
Internal hardware errors of the
detector
Consult with service engineers of
DRTECH
Consumption of a fully
charged battery pack is fast.
Performance decrease caused
by usage of long time.
Usage of battery pack in low
temperature environment.
Replace to new battery pack if the
battery pack has been used for a
long time. (Battery pack is a
consumable.)
Use battery pack in normal room
temperature environment. Charging
capacity of battery pack in a low
temperature environment will
decrease the capacity.
Battery pack or compartment
for installation of battery
pack is getting hot.
Battery pack failure.
Detector failure.
Do not use the battery pack.
Consult with service engineers of
DRTECH.