User's Manual

DRT-MAN-093
64/74
Mode
Description
AED/AWC Mode
3. The detector detects actual amount of X-rays without any connection to the
X-ray generator.
4. No signal used (No need of connector interface cable)
3.8.1 Lossless AED/AWC Mode
AED Mode is available for acquiring images without any connection to X-ray generator. Generator interface
cable is not required.
The acquired images are stored in the internal storage of the detector in order and the images can be
transmitted to the PC after connecting the detector to the PC. The transmitted images in the detector are
removed automatically.
Image processing and correction are available after connecting the detector to Viewer.
AWC(Auto Window Control) is a mode that automatically detects the end point of X-ray exposure and adjusts
the EWT(Exposure Window Time) set before the examination.
If an X-ray is detected within a shorter time than the preset maximum Exposure Window Time, the Exposure
Window is automatically closed and EWT is adjusted for faster operation of the detector.
Figure 9. AED Mode Configuration