User's Manual
DRT-MAN-093
27/74
Electromagnetic immunity
The Product is intended for use in the electromagnetic environment specified below. The user of the Product
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment – Guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
±(2, 4, 8) kV contact
±(2, 4, 8, 15) kV air
±(2, 4, 8) kV contact
±(2, 4, 8, 15) kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
(100kHz PRF)
±1 kV for input/output
lines(100kHz PRF)
±2 kV for power
supply lines
(100kHz PRF)
±1kV for input/output
lines(100kHz PRF)
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to ground
±1 kV line to line
±2 kV line to ground
Mains power quality should be that
of a typical commercial or hospital
environment
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0% U
T
for 0.5 cycles at
0˚, 45˚, 90˚, 135˚,
180˚, 225˚, 270˚, 315˚
0% U
T
for 1 cycles
40 % U
T
for 5 cycles
70% U
T
for 25/30
cycles
0% U
T
for 250/300
cycles
0% U
T
for 0.5 cycles
at 0˚, 45˚, 90˚, 135˚,
180˚, 225˚, 270˚,
315˚
0% U
T
for 1 cycles
40 % U
T
for 5 cycles
70% U
T
for 25/30
cycles
0% U
T
for 250/300
cycles
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the EVS
4343W / EVS 4343WG / EVS
4343WP / EVS 3643W / EVS
3643WG / EVS 3643 WP requires
continued operation during power
mains interruptions, it is
recommended that the EVS 4343W
/ EVS 4343WG / EVS 4343WP /
EVS 3643W / EVS 3643WG / EVS
3643WPbe powered from an
uninterruptible power supply
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment – Guidance
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of atypical location in a typical
commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150kHz-80MHz
6Vrms in ISM bands
between 0.15 MHz
and 80 MHz
Amateur 80% AM at 1
kHz
10 V/m
80 MHz to 2700 MHz
80% AM AT 1 kHz
3Vrms
150kHz-80MHz
6Vrms in ISM bands
between 0.15 MHz and
80 MHz
Amateur 80% AM at 1
kHz
10 V/m
80 MHz to 2700 MHz
80% AM AT 1 kHz
The RF communications
equipment should be used no
closer to any part of the Product
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separations
distance
d=2.0
√
P
80 MHz to 2.7 GHz
where P is the maximum output
power rating of the transmitter in