5 Acquire X-ray images and adjust the doses of radiation to match the Current Value to the Target Value
of Flat60. When the Current value reaches within 10% of Target Value, keep the doses of radiation at
that point.
6 Click the Get Flat60 button and acquire as many images as the number set at the top of the Stage. Or,
load the Flat60 image which is previously saved by clicking the Load Flat60 button.
7 Enter the detection range of Defect data into the Tolerance field based on the acquired images.
The recommended Tolerance value is 3%.
8 Click the Generate Defect Map button, create the new Defect Map Data file, and then save it.
No need to click the Load defect map button since the Defect Map file is automatically
created and loaded.