5.4.3.2 Auto defect detection
1 To create new defect map, apply the Defect data which is provided with the flat panel detector and then
click the Auto defect detection button.
2 Acquire X-ray images and adjust the doses of radiation to match the Current value to the Target Value of
Flat30. When the Current value reaches within 10% of Target Value, keep the doses of radiation at that
point.
For Flat30 and Flat60 the recommended Target Values are respectively 250 and 500.
3 To save the collected image data, check the Save acquired data checkbox. Once the images of Flat30
and Flat 60 are acquired, it is allowed to save as a file.
4 Click the Get Flat30 button to acquire as many images as the number set at the top of the Stage. Or,
load the Flat30 image which is previously saved by clicking the Load Flat30 button.