User's Manual
Wireless Service Manual
Page 91 of 124 RA14-11A-022
5.4.2 Offset Calibration
5.4.2.1 Pre-offset Calibration
1 Select Dark acquisition–Normal and click the Load pre-offset button to apply pre-offset data which is
provided with the flat panel detector.
2 When you create new pre-offset data, input the number of images to be acquired to Stage in the Pre-
offset calibration item and then click the Run pre-offset calibration button.
3 Acquiring image process will be proceeded automatically, then the average value of acquired images
will be displayed in Current Value and the current status will be displayed in Stage.
4 After acquiring images, a window for saving created pre-offset data appears, then specify a directory to
save the data.