Generator Communication Settings Remarks
CPI Serial Port COM 1
19200 bps / data 8 / stop 2 /
none / none
EMD Serial Port COM 3
19200 bps / data 8 / stop 1 /
none / none
Set WorkStation 2, 3
SEDECAL Shared memory Window title: window title For Human in DRApp.ini
[SETTINGS]
DRAppName = window title
For Vet in AppName.ini
[CONFIG]
APPNAME= window title
5.3 Detector Configuration
In this phase, defect pixels are corrected and gained pixels are calibrated using installed x-ray generator and
x-ray tube. The detector needs to warm up at least 30 minutes before performing the calibration.
The calibration should be performed on the following cases.
Detector installation
X-ray generator replacement
X-ray tube replacement
Exposure section Value change
Gain Type change