User's Manual
Wireless Service Manual
Page 40 of 124 RA14-11A-022
2.6.2 Timing of Signals (DR Trigger Mode)
A
A’
B B’
EXP_OK
EXP_REQ
Image Acquisition Time
Figure 2.12 Timing of Exposure signal
Image Acquisition Time+
Exposure request signal A (EXP_REQ) should be applied first, then exposure respond signal B
(EXP_OK).
Image Acquisition Time can be set from 40 ㎳ to 4,000 ㎳ with 1 ㎳ increment when default value
is 500 ㎳.
2.6.3 EXT_INF Port Pin Assignment
5 1
15
11
20 16
30
26
Figure 2.13 EXT_INF port pin assignments