2.6.1.2 AED Mode
Detector
X-ray System
Generator
Figure 2.11 AED Configuration
AED is available for acquiring images without any connection to X-ray generator. Generator interface
cable is not required.
Make sure to follow operating environmental requirements (Temp: +10 ~ +35).
If you use AED Mode out of operating environmental requirements, unwanted image
can be acquired without X-ray image acquiring.
Do not hit or drop the equipment. Unwanted images may be acquired in the AED
Mode if it receives strong jolt.
If you use a Grid under general imaging condition (Dose) or image a thick object in
the AED Mode, the efficiency of X-ray transformation may be reduced about 0% ~ 2%
compared to the DR Trigger Mode according to the thickness of the target.
If you image a thick object in the AED Mode with low X-ray tube voltage, an image
may not be acquired or horizontal line noise may occur.