User's Manual
Wireless Service Manual
Page 38 of 124 RA14-11A-022
2.6 X-ray Generator Interface
2.6.1 X-ray Exposure Mode
Mode Description
DR Trigger Mode 1 The detector receives EXP_REQ signal that X-ray generator is prepared to
generate X-rays.
2 The detector prepares image acquiring and then responds EXP_OK signal to the
X-ray generator.
3 The X-ray generator confirms EXP_OK signal and generates X-rays, then the
detector performs image acquiring according to Image Acquisition Time and
transmits the image data.
EXP_REQ (Generator → Detector), EXP_OK (Detector → Generator)
AED Mode The detector detects actual amount of X-rays without any connection to the X-ray
generator, and then performs image acquiring according to Image Acquisition Time
and transmits the image data.
No signal used (No need to connect Generator Interface Cable.)
Table 2.9 Exposure Mode
2.6.1.1 DR Trigger Mode
Detector
X-ray System
Generator
Exposure Request
Exposure OK
Generator Interface Cable
Figure 2.10 DR Trigger Mode Configuration
DR Trigger is the most common and recommended exposure mode. User can achieve the best quality
images with DR Trigger Mode.