Electromagnetic Immunity
The EUT is intended for use in the electromagnetic environment specified below.
The customer or user of the EUT should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 ㎸
contact
±8 ㎸ air
±6 ㎸
contact
±8 ㎸ air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2
㎸
for
power supply
lines
± 1 ㎸ for
input/output
lines
±2
㎸
for
power supply
lines
± 1 ㎸ for
input/output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 ㎸
differential
mode
±2 ㎸
common
mode
±1 ㎸
differential
mode
±2 ㎸
common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% Uт
(>95% dip in
Uт) for 0.5
cycle.
40% Uт
(60% dip in
Uт) for 5
cycles.
70% Uт
(30% dip in
Uт) for 25
cycles.
<5% Uт
(<95% dip in
Uт) for 5 sec.
<5% Uт
(>95% dip in
Uт) for 0.5
cycle.
40% Uт
(60% dip in
Uт ) for 5
cycles.
70% Uт
(30% dip in
Uт) for 25
cycles.
<5% Uт
(<95% dip in
Uт) for 5 sec.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
EUT image intensifier requires continued operation
during power mains interruptions, it is recommended
that the EUT image intensifier be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 ㎐)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE: Uт is the a.c. mains voltage prior to application of the test level.