Notes for Using the Equipment
System Diagnostic
The VXSetup software runs a system diagnostic. Run VXSetup software after installing the system and at
least once a year. If an error occurs, report the detailed error information to Vieworks local dealer or
distributor.
The owner is responsible for ensuring that the system diagnostic is performed every year.
Do not try to use the system if the system diagnostic is failed.
Calibration
To ensure optimal performance of the system, it is important to verify that the system is calibrated.
The owner is responsible for ensuring that the system calibration is performed after the
system installation is completed or the system is repaired. Do not try to use the system if
system calibration is not performed.
Distances measurements
Distances measurements in millimeters are possible only after distance calibration has been performed using
a reference object (refer to VXvue User Manual).
The operator is responsible for performing distance calibration with a reference object and
verifying the results of the distance calibration before taking any distance measurements
on an image.
Left/Right Marker
The operator is responsible for the correct and clear marking on the left or right side of the image to eliminate
possible errors.
The software includes an option to mark the image with L (left) or R (right) indicator from acquisition phase
through printing and archiving. If the operator chose, for any reason, not to use L/R markers, he must use an
alternative way to eliminate any possible mistake.