Page 115 of 124 RA14-11A-022
Test ITEM (9)
Image Acquisition Time
Frequency
D: Daily □ M: Monthly ■
Q: Quarterly □ A: Annually □
Objective:
Confirm the acquisition of time required to get image designed with optimal specifications.
Equipment:
Workstation, VXvue S/W
Power supply, Power Meter
X-ray GEN., X-ray Tube
Stop watch
Inspection Report Form
Procedure:
Turn on the power of the system.
Run VXvue.
Select Exposure Mode.
Register patient.
Expose X-rays with the condition of 50kVp and 5 mAs.
Check flat Image is acquired and displayed on monitor.
Performance and Corrective Action:
Check if the image is acquired in 5 to 7 seconds (including the processing time).
If image acquisition failed, check the X-ray is exposed normally and the triggering with X-ray generator is
properly configured.
If still image acquisition fail, service assistance is necessary.
Record result on Inspection Report Form.
Result:
Frequency: D: □, M: □, Q: □, A: □ Date: 20**-**-** Operator:
Limit of Acceptability:5~7 sec
Remarks: