User's Manual
Wireless Service Manual
Page 111 of 124 RA14-11A-022
Test ITEM (5)
PACS Server connection
Frequency
D: Daily ■ M: Monthly □
Q: Quarterly □ A: Annually □
Objective:
Confirm that VXvue is connected normally with PACS Server and can send performed study data to PACS
server.
Equipment:
Workstation, VXvue S/W, PACS Server
Power supply, Power Meter
Inspection Report Form
Procedure:
Turn on the power of the system.
Run VXvue.
Select Setting Mode.
Select DICOM tab.
Select Storage tab.
Select Storage Server.
Click the Echo button.
Performance and Corrective Action:
Check the connection status that is displayed in test result window.
If connection test failed, check PACS server is operating and configuration of PACS server connection is
correct.
If still PACS server connection fails, service assistance is necessary.
Record result on Inspection Report Form.
Result:
Frequency: D: □, M: □, Q: □, A: □ Date: 20**-**-** Operator:
Limit of Acceptability: N/A
Remarks: