User's Manual

Wireless Service Manual
Page 110 of 124 RA14-11A-022
Test ITEM (4)
Work list connection
Frequency
D: Daily M: Monthly
Q: Quarterly A: Annually
Objective:
Confirm that VXvue is connected normally with Worklist and can register study by querying from Worklist
server.
Equipment:
Workstation, VXvue S/W, Work List Server
Power supply, Power Meter
Inspection Report Form
Procedure:
Turn on the power of the system.
Run VXvue.
Select Setting Mode.
Select DICOM tab.
Select MWL tab.
Select Worklist server.
Click the Echo button.
Performance and Corrective Action:
Check the connection status that is displayed in test result window.
If connection test failed, check Worklist server is operating and configuration of Worklist server connection
is correct.
If still Worklist connection fails, service assistance is necessary.
Record result on Inspection Report Form.
Result:
Frequency: D: □, M: □, Q: □, A: □ Date: 20**-**-** Operator:
Limit of Acceptability: N/A
Remarks: