User's Manual
Wireless Service Manual
Page 108 of 124 RA14-11A-022
Test ITEM (2)
Temperature
Frequency
D: Daily ■ M: Monthly □
Q: Quarterly □ A: Annually □
Objective:
Check the monitoring in order to minimize the characteristic changes of Flat panel caused by external
temperature changes.
Equipment:
Workstation, VXvue S/W
Power supply, Power Meter
Inspection Report Form
Procedure:
Turn on the power of the system.
Run VXvue.
Check the temperature.
Performance and Corrective Action:
Confirm that the detector and ambient temperature is lower or higher than the operating temperature stated
in this Service Manual. If the detector and ambient temperature deviates from the operating range, adjust
the detector and ambient temperature properly to prevent poor image quality. If the image quality is reduced
after adjusting the temperature, service assistance is necessary.
Record result on Inspection Report Form.
Result:
Frequency: D: □, M: □, Q: □, A: □ Date: 20**-**-** Operator:
Limit of Acceptability: +10 ~ +35℃
Remarks: