Test Forms
Test ITEM (1)
Power Consumption
Frequency
D: Daily M: Monthly
Q: Quarterly A: Annually
Objective:
Confirm that the power operation of detector is normal.
Equipment:
Workstation, VXvue S/W
Power supply, Power Meter
Inspection Report Form,
Procedure:
Turn on the power of the system.
Run VXvue.
Read the data indicated by Power Meter.
Performance and Corrective Action:
Power Consumption of Max. 200VA is secured.
Light green colored LED should be turned on.
If the power consumption exceeds the tolerance limit, service assistance is necessary.
If the LED is not turned on, service assistance is necessary.
Record result on Inspection Report Form.
Result:
Frequency: D: □, M: □, Q: □, A: □ Date: 20**-**-** Operator:
Limit of Acceptability: Max. 200VA