5.4.4 Gain Pixel Correction
Before performing Gain Pixel Correction, consider followings:
Recommended SID is 150 (distance between X-ray tube and Detector).
Open the collimator of X-ray tube completely.
Align the center of the detector with the center of collimator.
Keep everything away from the detector surface.
1 Select Gain acquisition Normal in the left Step item. Expose X-rays and adjust dose of X-rays so
that Current Value reaches TargetValue. Keep the quantity of X-rays when Current Value remains within
10% difference of TargetValue.
Recommended TargetValue is 2000.
2 Click the Get button and expose X-rays ten times.
3 Save the Gain data to the desired folder.
4 Once all the procedures are completed, the OK button will be activated. Click the OK button to close the
dialog box.