User's Manual
Table Of Contents
- M-Turbo Ultrasound System User Guide
- Contents
- Introduction
- Chapter 1: Getting Started
- Chapter 2: System Setup
- Displaying the setup pages
- Restoring default settings
- A & B Key, Footswitch setup
- Administration setup
- Annotations setup
- Audio, Battery setup
- Cardiac Calculations setup
- Connectivity setup
- Date and Time setup
- Display Information setup
- IMT Calculations setup
- Network Status setup
- OB Calculations setup
- OB Custom Measurements setup
- OB Custom Tables setup
- Presets setup
- System Information setup
- USB Devices setup
- Chapter 3: Imaging
- Chapter 4: Measurements and Calculations
- Chapter 5: Troubleshooting and Maintenance
- Chapter 6: Safety
- Ergonomic safety
- Electrical safety classification
- Electrical safety
- Equipment safety
- Battery safety
- Clinical safety
- Hazardous materials
- Electromagnetic compatibility
- ALARA principle
- Acoustic artifacts
- Guidelines for reducing MI and TI
- Output display
- Transducer surface temperature rise
- Acoustic output measurement
- Acoustic output tables
- Labeling symbols
- Chapter 7: References
- Chapter 8: Specifications
- Glossary
- Index
Chapter 5: Troubleshooting and Maintenance 75
Troubleshooting
7 Airdry.
8 Examinethetransducerandcablefordamage
suchascracks,splitting,orfluidleaks.
Ifdamageisevident,discontinueuseofthe
transducer,andcontactSonoSiteoryourlocal
representative.
To clean and disinfect a transducer
(immersion method)
1 Disconnectthetransducerfromthesystem.
2 Removeanytransducersheath.
3 Cleanthesurfaceusing asoftclothlightly
dampenedinamildsoaporcompatible
cleaningsolutiontoremoveanyparticulate
matterorbodyfluids.
Applythesolutiontotheclothratherthanthe
surface.
4 Rinsewithwaterorwipewith
water‐dampenedcloth,andthenwipewitha
drycloth.
5 Mixthedisinfectantsolutioncompatiblewith
thetransducer,followingdisinfectantlabel
instructionsforsolutionstrengthsand
disinfectantcontactduration.
6 Immersethetransducerintothedisinfection
solutionnotmorethan12‐18 inches
(31‐46 cm)fromthepointwherethecable
enters
theconnector.
Followtheinstructionsonthedisinfectant
labelforthedurationofthetransducer
immersion.
7 Usingtheinstructionsonthedisinfectant
label,rinsetothepointoftheprevious
immersion,andthenairdryortoweldrywith
acleancloth.
8 Examinethetransducerandcablefordamage
suchas
cracks,splitting,orfluidleaks.
Ifdamageisevident,discontinueuseofthe
transducer,andcontactSonoSiteoryourlocal
representative.
Cleaning and disinfecting the battery
To clean and disinfect a battery (wipe
method)
1 Removethebatteryfromthesystem.
2 Cleanthesurfaceusingasoftclothlightly
dampenedinamildsoapordetergent
cleaningsolution.
Applythesolutiontotheclothratherthanthe
surface.
3 Wipethesurfaceswiththedisinfection
solution.Sani‐ClothHB,Sani‐ClothWipes,or
70%isopropylalcoholis
recommended.
4 Airdr.
Cleaning the footswitch
To clean the footswitch
1 Dampenanon‐abrasiveclothwithoneofthe
followingproducts:
•Isopropylalcohol
•Soapandwater
•Cidex
•SodiumHypochlorite5.25%(Bleach)
diluted10:1
2 Wringoutclothuntilslightlywetandthen
gentlyrubsoiledareauntilclean.
Caution:
To avoid damaging the battery, do
not allow cleaning solution or
disinfectant to come in contact
with the battery terminals.
Caution:
To avoid damaging the footswitch,
do not sterilize. It is not intended for
use in a sterile environment.