154 Temperature and humidity limits
•Powersupply
•SiteLinkImageManager
SonoCalcIMT
•SystemACpowercord(10ft/3.1m)
•TripleTransducerConnect
Peripherals
Seethemanufacturersspecificationsforthe
followingperipherals.
Medical grade
•Barcodescanner,serial
•Barcodescanner,USB
•Blackandwhiteprinter
Recommendedsourcesforprinterpaper:
ContactSonyat8006867669or
www.sony.com/professionaltoordersupplies
ortofindthelocaldistributor.
Colorprinter
•DVDrecorder
Non-medical grade
•Kensingtonsecuritycable
•USBstoragedevice
Temperature and humidity
limits
Note: Thetemperatu re,pressure,andhumiditylimits
applyonlytotheultrasoundsystem,transducers,and
battery.
Operating
System, battery, and transducer
10–40°C(50–104°F),15–95%R.H.
700to1060hPa(0.7to1.05ATM)
Shipping and storage
System and transducer
35–65°C(31–149°F),15–95%R.H.
500to1060hPa(0.5to1.05ATM)
Battery
20–60°C(4–140°F),15–95%R.H.(Forstorage
longerthan30days,storeatorbelowroom
temperature.)
500to1060hPa(0.5to1.05ATM)
Electrical
PowerSupplyInput:100240VAC,50/60Hz,2.0
AMax@100VAC
PowerSupplyOutput#1:15VDC,5.0AMax
PowerSupplyOutput#2:12VDC,2.3AMax
Battery
Thebatterycomprisessixlithiumioncellsplus
electronics,atemperaturesensor,andbattery
contacts.
Runtimeisuptotwo hours,depe ndingon
imagingmodeanddisplaybrightness.
Electromechanical safety
standards
EN606011:1997,EuropeanNorm,Medical
ElectricalEquipment–Part 1.General
RequirementsforSafety.
EN6060111:2001,EuropeanNorm,Medical
ElectricalEquipment–Part1.General
RequirementsforSafety–Section11.Collateral
Standard.SafetyRequirementsforMedical
ElectricalSystems.