User's Manual

Table Of Contents
Chapter 6: Safety 129
Safety
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#No data are reported for this operating condition since the global maximum index value is not reported for
the reason listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 29: Transducer Model: SLAx/13-6 Operating Mode: PW Doppler
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value (a) (a) 1.2 (b)
Associated Acoustic
Parameter
p
r.3
(MPa)
#
W
0
(mW)
—# 8.75 #
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm)
#
0.65
d
eq
(z
sp
)(cm)
0.13
f
c
(MHz)
#— # 6.00 #
Dim of A
aprt
X(cm) —# 0.24#
Y(cm)
—# 0.3 #
Other Information
PD (μsec) #
PRF (Hz) #
p
r
@PII
max
(MPa)
#
d
eq
@Pll
max
(cm)
0.13
Focal Length FL
x
(cm)
—# #
FL
y
(cm)
—# #
I
PA.3
@MI
max
(W/cm
2
)
#
Operating
Control
Conditions
Control 1: Exam Type Vas, Nrv,
Ven
Control 2: Sample Volume 5 mm
Control 3: PRF 20833
Control 4: Sample Vol. Position Zone 2