(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#No data are reported for this operating condition since the global maximum index value is not reported for
the reason listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 24: Transducer Model: P21x/5-1 Operating Mode: 2D
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.4 (a) 2.3
Associated Acoustic
Parameter
p
r.3
(MPa)
1.92
W
0
(mW)
#— —171.53
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm)
5.1
d
eq
(z
sp
)(cm)
f
c
(MHz)
1.93 # 1.94
Dim of A
aprt
X(cm) #— 1.9
Y(cm)
#— 1.3
Other Information
PD (μsec) 0.842
PRF (Hz) 4000
p
r
@PII
max
(MPa)
2.53
d
eq
@Pll
max
(cm)
Focal Length FL
x
(cm)
#— 18.46
FL
y
(cm)
#— 5.5
I
PA .3
@MI
max
(W/cm
2
)
355.1
Operating
Control
Conditions
Control 1: Exam Type Card Card
Control 2: Optimization Res Pen
Control 3: Depth 4.7/7.5
cm
27 cm
Control 4: THI On Off
Control 5: Sector Width Narrow Narrow