User's Manual
Table Of Contents
- M-Turbo Ultrasound System User Guide
- Contents
- Introduction
- Chapter 1: Getting Started
- Chapter 2: System Setup
- Displaying the setup pages
- Restoring default settings
- A & B Key, Footswitch setup
- Administration setup
- Annotations setup
- Audio, Battery setup
- Cardiac Calculations setup
- Connectivity setup
- Date and Time setup
- Display Information setup
- IMT Calculations setup
- Network Status setup
- OB Calculations setup
- OB Custom Measurements setup
- OB Custom Tables setup
- Presets setup
- System Information setup
- USB Devices setup
- Chapter 3: Imaging
- Chapter 4: Measurements and Calculations
- Chapter 5: Troubleshooting and Maintenance
- Chapter 6: Safety
- Ergonomic safety
- Electrical safety classification
- Electrical safety
- Equipment safety
- Battery safety
- Clinical safety
- Hazardous materials
- Electromagnetic compatibility
- ALARA principle
- Acoustic artifacts
- Guidelines for reducing MI and TI
- Output display
- Transducer surface temperature rise
- Acoustic output measurement
- Acoustic output tables
- Labeling symbols
- Chapter 7: References
- Chapter 8: Specifications
- Glossary
- Index
Chapter 6: Safety 123
Safety
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#No data are reported for this operating condition since the global maximum index value is not reported for
the reason listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 23: Transducer Model: P10x/8-4 Operating Mode: CW Doppler
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1A
aprt
>1
Global Maximum Index Value (a) — (a) — 2.1 2.0
Associated Acoustic
Parameter
p
r.3
(MPa)
#
W
0
(mW)
—# 40.72 30.00
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
z
bp
(cm)
—
z
sp
(cm)
#
0.7
d
eq
(z
sp
)(cm)
0.36
f
c
(MHz)
# — # — 4.00 4.00
Dim of A
aprt
X(cm) — # — 0.320 0.16
Y(cm)
—# — 0.70.7
Other Information
PD (μsec) #
PRF (Hz) #
p
r
@PII
max
(MPa)
#
d
eq
@Pll
max
(cm)
0.27
Focal Length FL
x
(cm)
—# — 0.92
FL
y
(cm)
—# — 5.0
I
PA.3
@MI
max
(W/cm
2
)
#
Operating
Control
Conditions
Control 1: Exam Type Card Card
Control 2: Depth Any Any
Control 3: Zone
Zone 3 Zone 0