(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#No data are reported for this operating condition since the global maximum index value is not reported for
the reason listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 22: Transducer Model: P10x/8-4 Operating Mode: PW Doppler
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.0 1.2 2.0 1.8
Associated Acoustic
Parameter
p
r.3
(MPa)
2.03
W
0
(mW)
36.25 34.4 31.5
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm)
2.1
0.8
d
eq
(z
sp
)(cm)
0.32
f
c
(MHz)
3.87 6.86 3.84 3.86
Dim of A
aprt
X(cm) 0.992 0.416 .224
Y(cm)
0.7 0.7 0.7
Other Information
PD (μsec) 1.28
PRF (Hz) 1563
p
r
@PII
max
(MPa)
2.70
d
eq
@Pll
max
(cm)
0.25
Focal Length FL
x
(cm)
—6.74 0.92
FL
y
(cm)
—5.0 5.0
I
PA.3
@MI
max
(W/cm
2
)
233
Operating
Control
Conditions
Control 1: Exam Type Crd Crd Neo Crd
Control 2: Sample Volume 1 mm 7 mm 12 mm 1 mm
Control 3: PRF/TDI 1563/
Off
Any/
On
15625/
Off
5208/
Off
Control 4: Sample Volume Position Zone 3 Zone 6 Zone 2 Zone 1