Chapter 6: Safety 111
Safety
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#No data are reported for this operating condition since the global maximum index value is not reported for
the reason listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 11: Transducer Model: C60x/5-2 Operating Mode: MMode
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.0 (a) (a) (b)
Associated Acoustic
Parameter
p
r.3
(MPa)
1.62
W
0
(mW)
—# ##
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm)
4.7
#
d
eq
(z
sp
)(cm)
#
f
c
(MHz)
2.85 # # #
Dim of A
aprt
X(cm) —# # #
Y(cm)
—# # #
Other Information
PD (μsec) 0.577
PRF (Hz) 800
p
r
@PII
max
(MPa)
2.576
d
eq
@Pll
max
(cm)
#
Focal Length FL
x
(cm)
—# #
FL
y
(cm)
—# #
I
PA.3
@MI
max
(W/cm
2
)
184.3
Operating
Control
Conditions
Control 1: Exam Type Any
Control 2: Optimization Pen
Control 3: Depth
7.8 cm
Control 4: MB (Multi Beam) Off or
On