User's Manual

Table Of Contents
102
Output display
ThesystemmeetstheAIUMoutputdisplaystandardforMIandTI(seelastreferencein
“Relatedguidancedocuments”below).Table 5indicatesforeachtransducerandoperating
modewheneithertheTIorMIisgreaterthanorequaltoavalueof1.0,thusrequiringdisplay.
Note: TheD2xtransducerhas
astaticcontinuouswave(CW)output.Thisoutputisfixed.Therefore,
TIandMIvaluescannotbechangedbyanysystemcontrolsavailabletotheuser.
Table 5: TI or MI 1.0
Transducer Model Index
2D/
MMode
CPD/
Color
PW
Doppler
CW
Doppler
C11x/8-5 MI No No No
TIC,TIB, or TIS No Yes Yes
C60x/5-2 MI Yes No No
TIC, TIB, or TIS No No Yes
D2x/2 MI ———No
TIC,TIB, or TIS Yes
HFL38x/13-6 MI No Yes Yes
TIC, TIB, or TIS No Yes Yes
ICTx/8-5 MI No No No
TIC, TIB, or TIS No No Yes
L25x/13-6 MI No Yes No
TIC,TIB, or TIS No No Yes
L38x/10-5 MI No Yes Yes
TIC, TIB, or TIS No Yes Yes
P10x/8-4 MI No Yes Yes No
TIC, TIB, or TIS Yes Yes Yes Yes
P21x/5-1 MI Yes Yes Yes No
TIC, TIB, or TIS Yes Yes Yes Yes
SLAx/13-6 MI No No No
TIC, TIB, or TIS No No Yes