Data Sheet
Page 44 ams Datasheet
Document Feedback [v1-04] 2018-Jul-09
AS7265x − Detailed Description
Control
ATINTTIME R/W
Set sensor
integration time.
Integration time =
<value> * ~2.8ms.
DEC 1-255 20
ATGAIN R/W
Set sensor gain: 0=1x
gain, 1=3.7x, 2=16x,
3=64x
DEC 0-3 1
ATINTRP R/W
Enable/Disable
interrupt pin
DEC
0 - Disable
1 - Enable Interrupt
pin functionality
0
AT TCSMD R/W
Set measurement
mode
DEC
0: Captures bank0 (1
integration period)
1: Captures bank1 (1
integration period)
2: Captures bank0+
bank1(2 integration
period)
3:Captures bank0+
bank1 in one shot
mode (2 integration
period)
2
ATINTRVL R/W
Set the sampling
interval as an integer
multiple of the
integration time. The
<value> is an integer
between [1...255].
A sampling
interval=1 implies a
sampling rate of 1x
the current
integration time.
A sampling
interval=255 implies
a slow sampling rate
of 255 times the
current integration
time
DEC 1…255 1
Commands Direction Description Format Value Range Default