Datasheet
Timing in HR Mode
The internal FIFO stores up to 32 samples, so that the
system processor does not need to read the data after
every sample. In HR mode (Figure 5), unlike in SpO
2
mode, temperature information is not necessary to interpret
the data. The user can select either the Red, IR, or Green
LED channel for heart rate.
Figure 5. Timing for Data Acquisition and Communication When in HR Mode
Table 16. Events Sequence for Figure 5 in HR Mode
EVENT DESCRIPTION COMMENTS
1 Enter into Mode
I
2
C Write Command sets MODE[2:0] = 0x02. Mask the A_FULL_EN
Interrupt.
2 FIFO is Almost Full, Interrupt Generated Interrupt is generated when the FIFO has only one empty space left.
3 FIFO Data is Read, Interrupt Cleared
4 Next Sample is Stored
New sample is stored at the new read pointer location. Eectively, it is
now the rst sample in the FIFO.
INT
I
2
C Bus
LED
OUTPUTS
IR
~
~
~
SAMPLE #1 SAMPLE #2 SAMPLE #3 SAMPLE #30 SAMPLE #31
1 2 3 4
15ms TO 300ms
IR IR IR IR IR IR
www.maximintegrated.com
Maxim Integrated
│
28
MAX30101 High-Sensitivity Pulse Oximeter and
Heart-Rate Sensor for Wearable Health