TELEMETRY TRANSMITTER (96281) OPERATIONS MANUAL A-2
A PPENDIX A — GUIDANCE AND MANUFACTURER S DECLARATION
Electromagnetic Immunity
Table 2—Electromagnetic Immunity
Immunity Test IEC 60601 Test Level
Compliance
Level
Electromagnetic
Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
N/A N/A
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
N/A N/A
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
for 0.5 cycle)
40% U
T
(60% dip in U
T
for 5 cycles)
70% U
T
(30% dip in U
T
for 25 cycles)
<5% U
T
(>95% dip in U
T
for 5 seconds)
N/A N/A
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m 3A/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
commercial or hospital
environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
D R A F T
22 June 2012