TELEMETRY TRANSMITTER (96281) OPERATIONS MANUAL 4-11
B ASIC OPERATIONS
Basic Modes of Operation
There are three basic operating modes to the telemetry transmitter.
1 Off
2 View
3 Non-View
Note:
The 96281-C automatically detects a sensor connection and activates or deactivates the SpO
2
option accordingly.
Table 1: Basic Modes of Operation
Operating Mode Description
Off
The device is off and is not in use if there are no batteries installed.
View
96281-B, 96281-C
1 User can view patient vital signs on the host monitor or
transmitter LCD.
2 User can initiate printing.
3 Patient vital signs are transmitted to the central station.
After 60 seconds of inactivity, the unit moves to the non-view mode.
Non-View
96281-A, 96281-B, 96281-C
Note:
Since the 96281-A does not have an LCD, it is either
continuously in non-view mode or is OFF.
1 The LCD is off on the 96281-B or 96281-C.
2 User can initiate printing.
3 Patient vital signs are transmitted to the central station.
4 User can return to the view mode by unlocking the user
interface (96281-B or 96281-C only).
To unlock the interface, push and hold
the up and down buttons
simultaneously for at least three
seconds.
D R A F T