TELEMETRY TRANSMITTER (96281) OPERATIONS MANUAL 3-17
ECG AND SP O
2
SENSOR DISCONNECTED — Check connection at adapter cable
This message and alarm indicate that the sensor is either disconnected or the wiring is faulty.
Check for proper sensor connection to the adapter cable.
If the message persists, replace the sensor and/or the adapter cable.
SENSOR OFF PATIENT — Check connection at patient
The telemetry transmitter does not detect a valid sensor input signal. Check the patient for proper
sensor placement.
The tissue between the LED and photodiode is too transmissive. If the sensor placement seems
correct and the message persists, try a sensor site with a thicker tissue bed.
Note:
This message is not available with all SpO
2
sensors.
INSUFFICIENT SIGNAL — Reposition or replace sensor
Insufficient signal for proper operation, indicated by a low deflection on the Sensorwatch signal
strength bar.
Poor sensor application or site. Correctly reapply or reposition to a better perfused site, or
massage the site.
If the message persists, then replace the sensor.
LOW SIGNAL STRENGTH — Reposition or replace sensor
When this message appears, the saturation and pulse rate continue to show. However, the
Sensorwatch bar flashes as an indication of a possible error condition.
Insufficient blood flow between the sensor light emitter and detector. Move the sensor to an area of
higher perfusion.
Poor sensor application. Reposition to place active components closer to the skin or locate to a
better perfused site.
The sensor site is below the blood pressure cuff. Move to another site.
If the message persists, then replace the sensor.
SPO2 SENSOR OFF PATIENT
SPO2 INSUFFICIENT SIGNAL
SPO2 AMBIENT LIGHT INTF.
SPO2 NOISY SIGNAL
SPO2 INSUFFICIENT SIGNAL