User's Manual
Table Of Contents
- Telemetry Transmitter
- Table of Contents
- Conventions Used in This Manual 1-1
- Nurses 1-7
- Monitor Technicians 1-7
- Biomedical Engineers 1-7
- Physicians 1-7
- Patients 1-7
- Sources of Interference 1-8
- Potential Sources of Damage 1-8
- Optional Leadwire Grouper 2-3
- Leadwire Color Codes 2-4
- Telemetry Channel Label 2-5
- Adult Electrode Placement 3-3
- Lead Fault Indication 3-4
- Noise Detection 3-4
- False Alarms 3-5
- Traditional Pulse Oximetry 3-5
- Electrodes, Leadwires, Sensors, and Sensor Cables 3-7
- Electrodes, Leadwires, Sensors and Sensor Cables 3-8
- Spacelabs Healthcare Technology 3-13
- Additional Information for Telemetry Products 3-13
- Telemetry 3-13
- Heart Rate Averaging 3-13
- Spacelabs Healthcare SpO2 Sensors 3-18
- Additional Information 3-18
- Transmitter Batteries 4-1
- Host Monitors 4-2
- Telemetry Receiver Module 4-2
- Assigning a Telemetry Channel 4-3
- Top, Front and Bottom View (96281-C) 4-4
- Rear View (96281-C) 4-5
- Front View (96281-A) 4-6
- Battery Compartment (96281-A, 96281-B, 96281-C) 4-7
- ECG 4-12
- SpO2 4-14
- Cleaning/Disinfecting 5-1
- Recommended Cleaning Solutions 5-2
- Basic Cleaning and Low-level Disinfection 5-3
- Cleaning ECG Leadwires 5-3
- Cleaning Buttons 5-3
- Cleaning the Battery Cover 5-3
- Table 1—Electromagnetic Emmissions A-1
- Table 2—Electromagnetic Immunity A-2
- Table 2—Electromagnetic Immunity (continued) A-3
- Table 3—Separation Distances A-4
- Introduction
- About the Transmitters
- ECG and SpO2
- ECG Overview
- Patient Preparation and Electrode Application
- To Set Up ECG Monitoring
- ECG Problem Solving
- SpO2 Overview
- Warnings and Cautions for SpO2
- Setting Up SpO2 Monitoring
- Ensuring Accurate SpO2 Monitoring
- SpO2 and Pulse Rate Specifications
- Using the Sensorwatch Feature
- Enabling and Adjusting Alarms
- Data Averaging
- Display Details at the Host Monitor
- Printing SpO2 Waveforms
- SpO2 Messages at the Host Monitor
- Sensors
- SpO2 Alarm Delays
- SpO2 Troubleshooting Guide
- Basic Operations
- Getting Started
- Basic Components
- Selecting Options for Leads
- Basic User Actions
- Basic Modes of Operation
- View Mode
- Status Messages at the Host Monitor
- Telemetry Transmitter with ECG Only Troubleshooting Guide
- Telemetry Transmitter with Display Troubleshooting Guide
- Telemetry Transmitter with Display and SpO2 Troubleshooting Guide
- Cleaning, Disinfecting, and Sterilization
- Appendix A — Guidance and Manufacturer’s Declaration
- Appendix B — Symbols

TELEMETRY TRANSMITTER (96281) OPERATIONS MANUAL 3-17
ECG AND SP O
2
SENSOR DISCONNECTED — Check connection at adapter cable
This message and alarm indicate that the sensor is either disconnected or the wiring is faulty.
• Check for proper sensor connection to the adapter cable.
• If the message persists, replace the sensor and/or the adapter cable.
SENSOR OFF PATIENT — Check connection at patient
• The telemetry transmitter does not detect a valid sensor input signal. Check the patient for proper
sensor placement.
• The tissue between the LED and photodiode is too transmissive. If the sensor placement seems
correct and the message persists, try a sensor site with a thicker tissue bed.
Note:
This message is not available with all SpO
2
sensors.
INSUFFICIENT SIGNAL — Reposition or replace sensor
• Insufficient signal for proper operation, indicated by a low deflection on the Sensorwatch signal
strength bar.
• Poor sensor application or site. Correctly reapply or reposition to a better perfused site, or
massage the site.
• If the message persists, then replace the sensor.
LOW SIGNAL STRENGTH — Reposition or replace sensor
When this message appears, the saturation and pulse rate continue to show. However, the
Sensorwatch bar flashes as an indication of a possible error condition.
• Insufficient blood flow between the sensor light emitter and detector. Move the sensor to an area of
higher perfusion.
• Poor sensor application. Reposition to place active components closer to the skin or locate to a
better perfused site.
• The sensor site is below the blood pressure cuff. Move to another site.
• If the message persists, then replace the sensor.
SPO2 SENSOR OFF PATIENT
SPO2 INSUFFICIENT SIGNAL
SPO2 AMBIENT LIGHT INTF.
SPO2 NOISY SIGNAL
SPO2 INSUFFICIENT SIGNAL
Table 5: Telemetry Message Equivalents (continued)
Telemetry SpO
2
Message
D R A F T
22 June 2012