User's Manual

Table Of Contents
A BOUT THE TRANSMITTERS
TELEMETRY TRANSMITTER (96281) OPERATIONS MANUAL 2-6
Lead Availability
ECG lead availability depends on which electrodes are connected to a transmitter, and whether the
transmitter is a Narrowband device (option N), or a Wideband device (option W).
Table 2 and Table 3 detail lead availability for narrowband and wideband transmitters, respectively,
based on which electrodes are connected.
Note:
Augmented leads aVR, aVL and aVF are only available at the monitor when leads I, II, and
III are all available.
For optimum performance, the right leg electrode should always be connected.
If an electrode is not connected, its lead fault indicators will flash on and off at the
transmitter.
If there are no leads available, all lead fault indicators will flash on and off at the transmitter.
The headings for each electrode list the AAMI and IEC lead designations (the IEC
designations are in parentheses).
If a lead is not available, the monitor will either show a blank or switch to an alternate
available lead.
Leads are only shown on a transmitter if the transmitter includes an optional display.
If a lead is not available, the transmitter will show a baseline trace.
Other combinations of disconnected electrodes yield no useful data, resulting in no lead
availability at either the monitor or the transmitter.
Table 2: Lead Availability with Narrowband transmitters (96281-xxxNx)
Connected Electrodes (X)
Available leads at the
monitor
Shown
leads at the
transmitter
RA
(R)
LA
(L)
LL
(F)
RL
(N)
V
(C)
X X X X X II, V I, II, III, V
X X X X II, V I, II, III, V
X X X X I, II, III, aVR, aVF, aVL I, II, III
X X X I, II, III, aVR, aVF, aVL I, II, III
X X X X III III
X X X III III
X X X III III
X X III III
X X X X II II
D R A F T
22 June 2012