User's Manual
SPECIFICATIONS
Ultraview
Digital
Telemetry
91341,
91343,
91347,
90478,
90479-A
Page 3 of 6
NIBP Display
(Refer to specifications for the 90217 ABP
Monitor)
Measurement Range (adult only) —
Systolic: 8.0 – 35.0 kPa
(60 – 260 mmHg)
Diastolic: 9.0 – 27.0 kPa
(30 – 200 mmHg)
Mean: 5.3 – 31.0 kPa
(40 – 230 mmHg)
Pressure Accuracy — ±2% or ±3 mmHg
(whichever is greater)
Resolution — 1 mmHg
Time Between Readings — selectable, from 6
to 120 minutes
Alarms — High and low alarms for all
measured parameters.
High range: 8.0 – 35.0 kPa
(60 – 260 mmHg)
Low range: 4.0 – 27.0 kPa
(30 – 200 mmHg)
TRANSMITTERS (91341, 91343, 90347)
ECG Transmission — View 2 of 7 available
leads from two vectors (91341) or four
vectors (91343, 91347) synchronized RF
digital signal.
Electrode Configuration — Individually
replaceable DIN standard safety lead wires.
Local lead fault indicators for each lead wire.
Multiparameter Transmission (90343) —
SpO
2
(saturation, SpO
2
sensor status, pulse
rate) and optional NIBP (systolic, diastolic,
mean pressure, measurement time, alarm
conditions) with the model 90217 ABP
monitor.
Additional Data Transmitted — Patient
record, low battery indicator, pacer flag,
patient ID code, and electrode connection
status.
Output Power — 1 mW ERP, typical
Spectral Efficiency — 0.11 bps/Hz
External Indicator — Yellow LED flashes
when battery level is low
Battery — 9 V battery; refer to Table 1 for
battery life expectancy
WMTS Frequency Band Option —
-05: 608-614 MHz
-09: 1395-1400 MHz
1427-1429.5 MHz
1429-1431.5 MHz
Transmitter Physical Dimensions
91343 (Multiparameter)
Height: 5.25 in (13.3 cm)
Width: 2.85 in (7.2 cm)
Depth: 1.18 in (2.9 cm)
Weight (w/out battery):
8.5 oz (241.0 gm)
91341/47 (ECG-only)
Height: 5.25 in (13.3 cm)
Width: 2.85 in (7.2 cm)
Depth: 0.98 in (2.5 cm)
Weight (w/out battery):
6.78 oz (192.7 gm)
MODULAR RECEIVER
Module Includes:
Module Configuration Manager capability
(refer to the Module Configuration Manager
chapter of the Ultraview Care Network
Operations Manual [P/N 070-1001-16] for
complete feature specifications).
Trends — (with appropriate mainframe option)
24 hours of trended data can be displayed in
1.5-, 3-, 6-, 12-, or 24-hour segments; data is
stored in 1-minute resolution.
High Level Analog Output —
ECG 1: Used for defibrillator
synchronization
Connector: 3-conductor TT phone jack
Dynamic Range: ±5 mV (±10%), rti
Gain: ECG x 1000 (±5%)
Bandwidth: 0.05 to 30 Hz
±10% (-3 dB)
Module Parameter Count — This module
counts as 1 or 2 parameters when computing
parameter capacity for monitors.
1 displayed ECG lead = 1 parameter.
2 displayed ECG leads = 2 parameters.
Receiver Sensitivity — Usable ECG signal to
-95 dBm