User's Manual
SPECIFICATIONS
Ultraview
Digital
Telemetry
91341,
91343,
91347,
90478,
90479-A
Page 2 of 6
PARAMETER CHARACTERISTICS
ECG
Maximum Input — ±4 mV ( ±10%)
DC Offset — Up to ±300 mV, with no more than
2% signal amplitude degradation
Overdrive Recovery Time — < 1 second
circuit settling time with offset voltage < 500
mV
Noise — < 30 µV p-v, refered to input (rti), at 30
Hz bandwidth
CMRR — > 85 dB (monitor mode)
QRS Detection — Detects QRS complexes
with durations of 40 to 120mS and
amplitudes of 0.2 to 4.0 mV (adult) or 0.15 to
4.0 mV (neonatal)
Defibrillator Protection — Meets IEC
60601-2-27, AAMI EC-13
Resolution — 2.5 µV per LSB, rti
Input Impedance — > 10 MΩ minimum
differential at 10 Hz
Gain Accuracy — ±5%
Pacer Rejection — Baseline shift < 0.2 mV
(measured at ECG x 1,000 output)
Pacer Detection — Detects pacer pulses of ±2
mV to ±700 mV with pulse widths of 0.2 to 2
msec and rise times 10% of width not to
exceed 100 µsec.
Signal Bandwidth — 0.05 to 30 Hz
±10% (-3 dB)
Sample Rate — 120 samples per second
SpO
2
SpO
2
Sensor Interface —
Red LED drive (max): 175 mA peak at 10%
duty cycle
IR LED drive (max): 105 mA peak at 10%
duty cycle
SpO
2
Measurement Method — Functional
saturation (oxygen saturation of functional
hemoglobin).
SpO
2
Measurement Mode — Continuous,
episodic (2 minutes, 5 minutes, and 30
minutes) sampling intervals; factory default
setting is continuous.
ST Segment Analysis
Resolution — 0.08 mm
Range — ±9 mm (1 mV = 10 mm)
Leads — ST Segment Analysis continously
performed on up to seven leads.
Alarms — Single lead or multiple leads;
individual leads can be deselected. Alarms
for absolute minimum and maximum ST
levels; changes in ST level over the last 5
minutes.
Displays — ST values: minimum/maximum/
current ST segment deviation and 5 minute
averaged segments for the last 30 minutes.
Measurement Points — Adjustable ST, PR,
and J Points
Trends — Up to 24 hours of trend data can be
displayed in 1.5-, 3-, 6-, 12-, or 24-hour time
tracks.
PATIENT MONITOR DISPLAY
CHARACTERISTICS
ECG Display
Heart Rate Range — 30 to 300 bpm; heart
rates > 300 bpm are displayed as “+++”.
Heart Rate Alarm Limits — High: 5 to 300
bpm, Low: 0 to 200 bpm; alarms
automatically enabled over a range of 40
(adult) or 100 (neonatal) to 300 bpm; heart
rates > 300 bpm are displayed as “+++”.
Accuracy — ±1% or ±2 beats per minute
(whichever is greater).
Numeric Update Rate — Every 3 seconds or
immediately at the onset of an alarm.
Trace Sweep Speeds — 50, 25, 12.5 mm/sec
Display Bandwidth — Two settiings: 0.5 to 30
Hz ±10% (-3 dB) in monitor mode, and 0.05
to 30 Hz ±25% (-3 dB at 50 mm per second)
in extended mode.
SpO
2
Display
Measurement Range — 30 to 100% O
2
Saturation
Saturation Accuracy — Sensor Dependent
Saturation Resolution — ±1%
Pulse Rate Range — 30 to 250 bpm
Pulse Rate Resolution — 1 bpm
Alarms — High and low saturation values;
factory default limits are: high 100%; low
85%.
High range: 31% – 100%
Low range: 30% – 99%
Numeric Update Rate — Every 2 seconds for
continuous SpO
2
readings.