User's Manual

071-0774-00 Rev. A 4-11
NIBP Troubleshooting Guide
Clinical Situation Possible Cause Solution
No NIBP displays Adapter cable not inserted correctly Remove and re-insert adapter cable
NIBP not enabled on 91343 or 90478 Enable NIBP function by setting transmitter
DIP switch 5 ON and setting DIP switch 8
OFF
90217 ABP not properly initialized Reinitialize 90217 ABP using 90121
91343 Low battery indicator constantly
illuminated
Call qualified service person
No NIBP readings can
be obtained
Incorrect or inoperative cuff in use Replace with cuff known to be operative
Tubing is kinked Locate kink and straighten tubing
Some arrhythmias (e.g., atrial
fibrillation and frequent ventricular
ectopy) may cause a single or repeated
failure to obtain a reading (may be due
to true beat-to-beat variations in
pressure)
Document arrhythmia if present, verify
pressure with another method, then follow
hospital procedure for care of this type of
patient
Excessive patient motion or muscle
contractions associated with shivering
or severe pain
Ensure that patient is quiet with minimal
movement during NIBP readings; minimize
patient’s shivering
Blood pressure outside of
measurement range
Verify extremely high or low pressure with
another method
Intermittent or
complete failure to
operate
90217 ABP error Remove 90217 ABP from service; record
event code; and call qualified service
person
Apparent incorrect
value
Wrong size cuff for patient Measure patient’s limbs at the midpoint;
match limb measurement to range specified
on cuff (undersizing the cuff results in the
greatest degree of error)
Cuff is damaged Replace with good cuff
Excessive patient motion, shivering or
severe pain
Ensure patient is quiet with minimal
movement during NIBP readings; minimize
patient’s shivering
False high readings may be the result
of venous congestion caused by
frequent readings
Reduce frequency of readings
Cuff too loose or positioned incorrectly Tighten cuff or reposition appropriately