Troubleshooting Guide
Clinical Situation Possible Cause Solution
No SpO
2
label is
displayed
SpO
2
is not enabled at the 91343 Be sure transmitter DIP switch 1 and 2
are set correctly
SpO
2
is not enabled at the 90478
receiver
Be sure transmitter DIP switch 8 is
OFF
SpO
2
value displays ??? Sensor not connected to patient Re-attach sensor
Adapter cable not connected to module
properly
Correctly connect the adapter cable
Sensor not connected to adapter cable Correctly connect the sensor
Excessive patient motion Urge patient to remain still while
reading is in progress
Transmitter is in the initialization phase
(the first 15 seconds after sensor
application)
Wait until initialization is complete
Low battery indicator constantly
illuminated
Call qualified service person
Insufficient signal or
noisy signal
Sensor placement not optimum Move sensor to a site which has better
perfusion
Align LED with sensor photodetector
Sensor placed below blood pressure
cuff
Move sensor to an alternate limb
Intermittent or complete
failure to operate
Depleted battery Replace battery
Low battery light constantly illuminated Call qualified service person
Factors that cause
significant variances in
sensor accuracy
Presence of dysfunctional hemoglobins
(COHb, MetHb)
Follow hospital procedure for
determining oxygenation in these
patients
Presence of intravascular dyes
(indocyamine green, methylene blue)
depending on their concentration in the
blood stream
Follow hospital procedure for
determining oxygenation in these
patients
High ambient light level Reduce light levels near patient; wrap
sensor with light blocking material
Electrosurgical interference Ultraview digital telemetry is contra-
indicated for electrosurgical use
Patient is significantly anemic (Hb less
than 5 gm/dl) or patient has received
large amounts of IV solutions
Follow hospital procedure for
determining oxygenation in these
patients