User's Manual
Table Of Contents
- Directory of Keys - UCW and Ultraview 1700
- Ultraview Digital Telemetry
- General Telemetry Overview
- Cleaning
- Assigning a Telemetry Channel
- Tuning a Receiver for a Bedside
- Entering Patient Information
- Discharging a Patient
- Acknowledging Signal Loss
- Setting Battery Status Alarms
- Controlling Patient-Initiated Recordings
- Telemetry Alarm Message Summary
- ECG Overview
- Setting Up ECG Monitoring
- Display Detail
- Monitoring Paced ECG Patients
- Restoring Default Settings
- Changing the Display Resolution
- Selecting Options for Lead Display
- ECG Alarm Message Summary
- SpO2 Overview (90343 only)
- Setting Up SpO2 Monitoring
- Ensuring Accurate Monitoring
- Setting or Adjusting Alarm Limits
- Setting SpO2 Data Averaging Period and Sampling Interval
- Viewing Pulse Rate
- SpO2 with Intra-Aortic Balloon Pumps
- Using SpO2 with Neonates
- SpO2 Alarm Message Summary
- NIBP Overview (90343 only)
- Setting Up NIBP Monitoring
- Setting Up the ABP Monitor
- Setting or Adjusting Alarm Limits
- Displaying New or Previous Readings
- NIBP Alarm Message Summary
- Alarm Message Summary
- Accessories

NIBP Troubleshooting Guide (continued)
Clinical Situation Possible Cause Solution
Variable readings occur
■ Some arrhythmias may cause beat-to-
beat pressure and NIBP readings
■ Document arrhythmia if present, verify
pressure using another method, then
follow hospital procedure for care of this
type of patient
■ Larger than normal influence of
respiratory phases on blood pressure
(inspiratory fall in blood pressure;
expiratory rise)
■ NIBP software usually compensates for
normal variation
No NIBP readings or
questionable values in the
presence of shock
■ Peripheral vascular changes
experienced during shock may reduce
the reliability of blood pressure
readings obtained with any indirect
method; peripheral pulses may be
diminished or absent
■ Consider invasive pressure
measurements in patients with symptoms
of shock or in any patient who rapidly
becomes unstable for unknown reasons
90217 displays “LLL” and
alarm sounds
■ Low main battery condition ■ Turn off and replace batteries within 60
seconds after removal to continue
monitoring.
Cuff too tight
■ Cuff placed on patient too tightly ■ Reposition the cuff
■ Air pump staying on too long ■ Return unit to Spacelabs Medical for
service
Cuff too loose
■ Cuff placed on patient too loosely ■ Reposition the cuff
■ Air pump not staying on long enough ■ Return unit to Spacelabs Medical for
service