User's Manual

NIBP Troubleshooting Guide (continued)
Clinical Situation Possible Cause Solution
Variable readings occur
Some arrhythmias may cause beat-to-
beat pressure and NIBP readings
Document arrhythmia if present, verify
pressure using another method, then
follow hospital procedure for care of this
type of patient
Larger than normal influence of
respiratory phases on blood pressure
(inspiratory fall in blood pressure;
expiratory rise)
NIBP software usually compensates for
normal variation
No NIBP readings or
questionable values in the
presence of shock
Peripheral vascular changes
experienced during shock may reduce
the reliability of blood pressure
readings obtained with any indirect
method; peripheral pulses may be
diminished or absent
Consider invasive pressure
measurements in patients with symptoms
of shock or in any patient who rapidly
becomes unstable for unknown reasons
90217 displays LLL and
alarm sounds
Low main battery condition Turn off and replace batteries within 60
seconds after removal to continue
monitoring.
Cuff too tight
Cuff placed on patient too tightly Reposition the cuff
Air pump staying on too long Return unit to Spacelabs Medical for
service
Cuff too loose
Cuff placed on patient too loosely Reposition the cuff
Air pump not staying on long enough Return unit to Spacelabs Medical for
service