User's Manual

NIBP Troubleshooting Guide
Clinical Situation Possible Cause Solution
No NIBP displays
Adapter cable not inserted correctly Remove and re-insert adapter cable
NIBP not enabled on 90343 or 90478 Enable NIBP function by setting
transmitter DIP switch 5 ON and setting
DIP switch 8 OFF
90217 ABP not properly initialized Reinitialize 90217 ABP using 90121
90343 Low battery indicator constantly
illuminated
Call qualified service person
No NIBP readings can be
obtained
Incorrect or inoperative cuff in use Replace with cuff known to be operative
Tubing is kinked Locate kink and straighten tubing
Some arrhythmias (e.g., atrial
fibrillation and frequent ventricular
ectopy) may cause a single or repeated
failure to obtain a reading (may be due
to true beat-to-beat variations in
pressure)
Document arrhythmia if present, verify
pressure with another method, then follow
hospital procedure for care of this type of
patient
Excessive patient motion or muscle
contractions associated with shivering
or severe pain
Ensure that patient is quiet with minimal
movement during NIBP readings;
minimize patients shivering
Blood pressure outside of
measurement range
Verify extremely high or low pressure with
another method
Intermittent or complete
failure to operate
90217 ABP error Remove 90217 ABP from service; record
event code; and call qualified service
person
Apparent incorrect value
Wrong size cuff for patient Measure patients limbs at the midpoint;
match limb measurement to range
specified on cuff (undersizing the cuff
results in the greatest degree of error)
Cuff is damaged Replace with good cuff
Excessive patient motion, shivering or
severe pain
Ensure patient is quiet with minimal
movement during NIBP readings;
minimize patients shivering
False high readings may be the result
of venous congestion caused by
frequent readings
Reduce frequency of readings
Cuff too loose or positioned incorrectly Tighten cuff or reposition appropriately
90217 ABP Display is
incorrect
Data not retained Replace backup battery
Low or no power Check the batteries for a full charge; if
needed, replace or recharge the batteries
May be one of the following: time-out,
no reading due to air leak in the
system, improper cuff size, cuff size not
properly attached to the 90217 ABP
Isolate cause and correct