User's Manual
Table Of Contents
- Directory of Keys - UCW and Ultraview 1700
- Ultraview Digital Telemetry
- General Telemetry Overview
- Cleaning
- Assigning a Telemetry Channel
- Tuning a Receiver for a Bedside
- Entering Patient Information
- Discharging a Patient
- Acknowledging Signal Loss
- Setting Battery Status Alarms
- Controlling Patient-Initiated Recordings
- Telemetry Alarm Message Summary
- ECG Overview
- Setting Up ECG Monitoring
- Display Detail
- Monitoring Paced ECG Patients
- Restoring Default Settings
- Changing the Display Resolution
- Selecting Options for Lead Display
- ECG Alarm Message Summary
- SpO2 Overview (90343 only)
- Setting Up SpO2 Monitoring
- Ensuring Accurate Monitoring
- Setting or Adjusting Alarm Limits
- Setting SpO2 Data Averaging Period and Sampling Interval
- Viewing Pulse Rate
- SpO2 with Intra-Aortic Balloon Pumps
- Using SpO2 with Neonates
- SpO2 Alarm Message Summary
- NIBP Overview (90343 only)
- Setting Up NIBP Monitoring
- Setting Up the ABP Monitor
- Setting or Adjusting Alarm Limits
- Displaying New or Previous Readings
- NIBP Alarm Message Summary
- Alarm Message Summary
- Accessories

NIBP Troubleshooting Guide
Clinical Situation Possible Cause Solution
No NIBP displays
■ Adapter cable not inserted correctly ■ Remove and re-insert adapter cable
■ NIBP not enabled on 90343 or 90478 ■ Enable NIBP function by setting
transmitter DIP switch 5 ON and setting
DIP switch 8 OFF
■ 90217 ABP not properly initialized ■ Reinitialize 90217 ABP using 90121
■ 90343 Low battery indicator constantly
illuminated
■ Call qualified service person
No NIBP readings can be
obtained
■ Incorrect or inoperative cuff in use ■ Replace with cuff known to be operative
■ Tubing is kinked ■ Locate kink and straighten tubing
■ Some arrhythmias (e.g., atrial
fibrillation and frequent ventricular
ectopy) may cause a single or repeated
failure to obtain a reading (may be due
to true beat-to-beat variations in
pressure)
■ Document arrhythmia if present, verify
pressure with another method, then follow
hospital procedure for care of this type of
patient
■ Excessive patient motion or muscle
contractions associated with shivering
or severe pain
■ Ensure that patient is quiet with minimal
movement during NIBP readings;
minimize patient’s shivering
■ Blood pressure outside of
measurement range
■ Verify extremely high or low pressure with
another method
Intermittent or complete
failure to operate
■ 90217 ABP error ■ Remove 90217 ABP from service; record
event code; and call qualified service
person
Apparent incorrect value
■ Wrong size cuff for patient ■ Measure patient’s limbs at the midpoint;
match limb measurement to range
specified on cuff (undersizing the cuff
results in the greatest degree of error)
■ Cuff is damaged ■ Replace with good cuff
■ Excessive patient motion, shivering or
severe pain
■ Ensure patient is quiet with minimal
movement during NIBP readings;
minimize patient’s shivering
■ False high readings may be the result
of venous congestion caused by
frequent readings
■ Reduce frequency of readings
■ Cuff too loose or positioned incorrectly ■ Tighten cuff or reposition appropriately
90217 ABP Display is
incorrect
■ Data not retained ■ Replace backup battery
■ Low or no power ■ Check the batteries for a full charge; if
needed, replace or recharge the batteries
■ May be one of the following: time-out,
no reading due to air leak in the
system, improper cuff size, cuff size not
properly attached to the 90217 ABP
■ Isolate cause and correct