User's Manual

Ultraview Digital Telemetry
37
split screen central monitor appear in a format slightly different from that of
bedside or full screen central monitors. Depending on the patient monitors display
size, the title NIBP may not appear.
Figure Telemetry-8: Display Zone Full Screen
Figure Telemetry-9: Display Zone Split Screen
Last systolic reading
Last diastolic reading
Last mean reading
Hour and minutes of last reading
Equal sign becomes bell symbol when NIBP alarms are enabled
Setting or Adjusting Alarm Limits
You can define pressure alarm limits for systolic, diastolic, and mean values. The
default setting for alarms is OFF. Refer to the Alarms chapter in this manual for
Ultraview system alarm functions. The alarm limits defaults are listed in Table
Telemetry-6: Alarm Limits.
Table Telemetry-6: Alarm Limits
Refer to the Module Configuration Manager chapter for NIBP parameter tables
that list available user settings and factory defaults for this parameter.
High Low
systolic 180 mmHg (24.0 kPa) 100 mmHg (13.3 kPa)
diastolic 120 mmHg (16.0 kPa) 60 mmHg (8.0 kPa)
mean 130 mmHg (17.3 kPa) 80 mmHg (12.0 kPa)
E
C
G
XX YYY
PACED
A=XX
A/M 10
ROW 5
130
40
78
CHAN xxxx
ECG WAVEFORM ZONE
BED NAME PATIENT NAME
NIBP=120/68(94) hh:mm SpO2=98% hh:mm
E
C
G
HR=XXX A=XX LEAD
ECG WAVEFORM ZONE
NIBP=120/68(94) hh:mm SpO2=98% hh:mm
CHAN xxxx
BED NAME PATIENT NAME
To set or adjust NIBP alarms:
1 Touch ECG
2 Touch ALARM LIMITS
3 Touch NIBP ALARM LIMITS
4 Select NIBP ALM ON
5 Select SYS, DIA, or MEAN
6 Select HI= or LO=
7 Use arrow keys to adjust