User's Manual
Table Of Contents
- Directory of Keys - UCW and Ultraview 1700
- Ultraview Digital Telemetry
- General Telemetry Overview
- Cleaning
- Assigning a Telemetry Channel
- Tuning a Receiver for a Bedside
- Entering Patient Information
- Discharging a Patient
- Acknowledging Signal Loss
- Setting Battery Status Alarms
- Controlling Patient-Initiated Recordings
- Telemetry Alarm Message Summary
- ECG Overview
- Setting Up ECG Monitoring
- Display Detail
- Monitoring Paced ECG Patients
- Restoring Default Settings
- Changing the Display Resolution
- Selecting Options for Lead Display
- ECG Alarm Message Summary
- SpO2 Overview (90343 only)
- Setting Up SpO2 Monitoring
- Ensuring Accurate Monitoring
- Setting or Adjusting Alarm Limits
- Setting SpO2 Data Averaging Period and Sampling Interval
- Viewing Pulse Rate
- SpO2 with Intra-Aortic Balloon Pumps
- Using SpO2 with Neonates
- SpO2 Alarm Message Summary
- NIBP Overview (90343 only)
- Setting Up NIBP Monitoring
- Setting Up the ABP Monitor
- Setting or Adjusting Alarm Limits
- Displaying New or Previous Readings
- NIBP Alarm Message Summary
- Alarm Message Summary
- Accessories

Ultraview Digital Telemetry
37
split screen central monitor appear in a format slightly different from that of
bedside or full screen central monitors. Depending on the patient monitor’s display
size, the title “NIBP” may not appear.
Figure Telemetry-8: Display Zone — Full Screen
Figure Telemetry-9: Display Zone — Split Screen
➊ Last systolic reading
➋ Last diastolic reading
➌ Last mean reading
➍ Hour and minutes of last reading
➎ Equal sign becomes bell symbol when NIBP alarms are enabled
Setting or Adjusting Alarm Limits
You can define pressure alarm limits for systolic, diastolic, and mean values. The
default setting for alarms is OFF. Refer to the Alarms chapter in this manual for
Ultraview system alarm functions. The alarm limits defaults are listed in Table
Telemetry-6: Alarm Limits.
Table Telemetry-6: Alarm Limits
Refer to the Module Configuration Manager chapter for NIBP parameter tables
that list available user settings and factory defaults for this parameter.
High Low
systolic 180 mmHg (24.0 kPa) 100 mmHg (13.3 kPa)
diastolic 120 mmHg (16.0 kPa) 60 mmHg (8.0 kPa)
mean 130 mmHg (17.3 kPa) 80 mmHg (12.0 kPa)
E
C
G
XX YYY
PACED
A=XX
A/M 10
ROW 5
130
40
78
CHAN xxxx
ECG WAVEFORM ZONE
BED NAME PATIENT NAME
NIBP=120/68(94) hh:mm SpO2=98% hh:mm
➌
➍
➊
➋
➎
E
C
G
HR=XXX A=XX LEAD
ECG WAVEFORM ZONE
NIBP=120/68(94) hh:mm SpO2=98% hh:mm
CHAN xxxx
BED NAME PATIENT NAME
➌
➍
➋
➊
➎
To set or adjust NIBP alarms:
1 Touch ECG
2 Touch ALARM LIMITS
3 Touch NIBP ALARM LIMITS
4 Select NIBP ALM ON
5 Select SYS, DIA, or MEAN
6 Select HI= or LO=
7 Use arrow keys to adjust