User's Manual
Table Of Contents
- Directory of Keys - UCW and Ultraview 1700
- Ultraview Digital Telemetry
- General Telemetry Overview
- Cleaning
- Assigning a Telemetry Channel
- Tuning a Receiver for a Bedside
- Entering Patient Information
- Discharging a Patient
- Acknowledging Signal Loss
- Setting Battery Status Alarms
- Controlling Patient-Initiated Recordings
- Telemetry Alarm Message Summary
- ECG Overview
- Setting Up ECG Monitoring
- Display Detail
- Monitoring Paced ECG Patients
- Restoring Default Settings
- Changing the Display Resolution
- Selecting Options for Lead Display
- ECG Alarm Message Summary
- SpO2 Overview (90343 only)
- Setting Up SpO2 Monitoring
- Ensuring Accurate Monitoring
- Setting or Adjusting Alarm Limits
- Setting SpO2 Data Averaging Period and Sampling Interval
- Viewing Pulse Rate
- SpO2 with Intra-Aortic Balloon Pumps
- Using SpO2 with Neonates
- SpO2 Alarm Message Summary
- NIBP Overview (90343 only)
- Setting Up NIBP Monitoring
- Setting Up the ABP Monitor
- Setting or Adjusting Alarm Limits
- Displaying New or Previous Readings
- NIBP Alarm Message Summary
- Alarm Message Summary
- Accessories

SpO
2
Troubleshooting Guide
Clinical Situation Possible Cause Solution
No SpO
2
label is
displayed
■ SpO
2
is not enabled at the 90343 ■ Be sure transmitter DIP switch 1 and 2
are set correctly
■ SpO
2
is not enabled at the 90478
receiver
■ Be sure transmitter DIP switch 8 is
OFF
SpO
2
value displays ??? ■ Sensor not connected to patient ■ Re-attach sensor
■ Adapter cable not connected to module
properly
■ Correctly connect the adapter cable
■ Sensor not connected to adapter cable ■ Correctly connect the sensor
■ Excessive patient motion ■ Urge patient to remain still while
reading is in progress
■ Transmitter is in the initialization phase
(the first 15 seconds after sensor
application)
■ Wait until initialization is complete
■ Low battery indicator constantly
illuminated
■ Call qualified service person
Insufficient signal or
noisy signal
■ Sensor placement not optimum ■ Move sensor to a site which has better
perfusion
■ Align LED with sensor photodetector
■ Sensor placed below blood pressure cuff ■ Move sensor to an alternate limb
Intermittent or complete
failure to operate
■ Depleted battery ■ Replace battery
■ Low battery light constantly illuminated ■ Call qualified service person
Factors that cause
significant variances in
sensor accuracy
■ Presence of dysfunctional hemoglobins
(COHb, MetHb)
■ Follow hospital procedure for
determining oxygenation in these
patients
■ Presence of intravascular dyes
(indocyamine green, methylene blue)
depending on their concentration in the
blood stream
■ Follow hospital procedure for
determining oxygenation in these
patients
■ High ambient light level ■ Reduce light levels near patient; wrap
sensor with light blocking material
■ Electrosurgical interference ■ Ultraview digital telemetry is contra-
indicated for electrosurgical use
■ Patient is significantly anemic (Hb less
than 5 gm/dl) or patient has received
large amounts of IV solutions
■ Follow hospital procedure for
determining oxygenation in these
patients