User's Manual
Table Of Contents
- Directory of Keys - UCW and Ultraview 1700
- Ultraview Digital Telemetry
- General Telemetry Overview
- Cleaning
- Assigning a Telemetry Channel
- Tuning a Receiver for a Bedside
- Entering Patient Information
- Discharging a Patient
- Acknowledging Signal Loss
- Setting Battery Status Alarms
- Controlling Patient-Initiated Recordings
- Telemetry Alarm Message Summary
- ECG Overview
- Setting Up ECG Monitoring
- Display Detail
- Monitoring Paced ECG Patients
- Restoring Default Settings
- Changing the Display Resolution
- Selecting Options for Lead Display
- ECG Alarm Message Summary
- SpO2 Overview (90343 only)
- Setting Up SpO2 Monitoring
- Ensuring Accurate Monitoring
- Setting or Adjusting Alarm Limits
- Setting SpO2 Data Averaging Period and Sampling Interval
- Viewing Pulse Rate
- SpO2 with Intra-Aortic Balloon Pumps
- Using SpO2 with Neonates
- SpO2 Alarm Message Summary
- NIBP Overview (90343 only)
- Setting Up NIBP Monitoring
- Setting Up the ABP Monitor
- Setting or Adjusting Alarm Limits
- Displaying New or Previous Readings
- NIBP Alarm Message Summary
- Alarm Message Summary
- Accessories

Ultraview Care Network
26
Setting SpO
2
Data Averaging Period and
Sampling Interval
SpO
2
data averaging is used to smooth the oximetry saturation value by
averaging the patient input values over 4, 8, or 16 seconds. This selection is made
by setting the DIP switches 1 and 2 beneath the battery compartment in the 90343
digital telemetry multi-parameter transmitter. The default value is 8 seconds.
Refer to Figure Telemetry-5: DIP Switch Setting in Battery Compartment.
Table Telemetry-3: DIP Switch 1 and 2 Settings
The current setting of the SpO
2
averaging period may be displayed by pressing
the ECG CHANNEL FORMAT key and enabling SpO
2
.
Figure Telemetry-5: DIP Switch Setting in Battery Compartment
The sampling interval selection enables you to determine how often an SpO
2
measurement will be taken. Less frequent SpO
2
readings can extend the usable
life of the battery. (Refer to the Ultraview Digital Telemetry Products data sheet,
!
• Setting both DIP switches 1 and 2 to ON disables SpO
2
data
transmission.
• To enable SpO
2
, remove the battery, set the selected interval,
and re-install the battery.
• Disabling SpO
2
operation in the 90343 transmitter lengthens
battery life.
C
AUTI
O
N:
• Use care when configuring the DIP switches. Avoid using
pens or pencils to configure the DIP switches since they
may cause contamination. Avoid using sharp cutting
instruments which may cause physical damage.
DIP Switch 1 DIP Switch 2 Effect
OFF OFF 4 seconds averaging enabled
OFF ON 8 seconds averaging enabled (default)
ON OFF 16 seconds averaging enabled
ON ON Disable SpO
2
operation
To set SpO
2
data averaging
period and sampling interval, set
transmitter DIP switches 1
through 4 to correct
configuration
1 2 3 4
SpO
2
averaging
period
SpO
2
reading
interval
5 6 7 8
ON
OFF
enable
normal
enabledisable
IABP
disable
NIBP
Adult
enable
NIBP
enable
IABP
enable
Neonate
enable
Service use
operation