User's Manual
Table Of Contents
- Directory of Keys - UCW and Ultraview 1700
- Ultraview Digital Telemetry
- General Telemetry Overview
- Cleaning
- Assigning a Telemetry Channel
- Tuning a Receiver for a Bedside
- Entering Patient Information
- Discharging a Patient
- Acknowledging Signal Loss
- Setting Battery Status Alarms
- Controlling Patient-Initiated Recordings
- Telemetry Alarm Message Summary
- ECG Overview
- Setting Up ECG Monitoring
- Display Detail
- Monitoring Paced ECG Patients
- Restoring Default Settings
- Changing the Display Resolution
- Selecting Options for Lead Display
- ECG Alarm Message Summary
- SpO2 Overview (90343 only)
- Setting Up SpO2 Monitoring
- Ensuring Accurate Monitoring
- Setting or Adjusting Alarm Limits
- Setting SpO2 Data Averaging Period and Sampling Interval
- Viewing Pulse Rate
- SpO2 with Intra-Aortic Balloon Pumps
- Using SpO2 with Neonates
- SpO2 Alarm Message Summary
- NIBP Overview (90343 only)
- Setting Up NIBP Monitoring
- Setting Up the ABP Monitor
- Setting or Adjusting Alarm Limits
- Displaying New or Previous Readings
- NIBP Alarm Message Summary
- Alarm Message Summary
- Accessories

Ultraview Digital Telemetry
25
MSG ALM DELAY Key
This key sets the number of seconds the system will wait before it issues an alarm
tone following any of the following messages:
• SpO2 UNAVAILABLE
• SpO2 FAULTY SENSOR
• SpO2 SENSOR DISCONNECTED
• SpO2 SENSOR OFF PATIENT
• SpO2 INSUFFICIENT SIGNAL
• SpO2 AMBIENT LIGHT INTF.
• SpO2 NOISY SIGNAL
When this feature is OFF, the key label will read “MSG ALM DELAY OFF”. When
it is ON, the label will read “MSG ALM DELAY xx” where “xx” is the value, in
seconds, of the delay.
To set the message delay time:
1. Touch MSG ALM DELAY xx (or MSG ALM DELAY OFF).
2. Touch the up and down arrow keys until the value is set as desired. Possible
settings are OFF, 10, 20, 30, 40, 50, or 60 seconds.
!
• If you press the down arrow key after the lowest value has been
reached, the following message will appear on the prompt line:
Minimum message alarm delay time has been reached.
• If you press the up arrow key after the highest value has been
reached, the following message will appear on the prompt line:
Maximum message alarm delay time has been reached.