User's Manual

Ultraview Digital Telemetry
23
All sensors require an adapter cable between the sensor and the transmitter.
Because the adaptor cable is reusable, do not discard it when you have finished
using a disposable oximetry sensor. Disconnect the sensor cable from the adapter
cable before discarding the sensor.
To connect the SpO
2
adapter cable to the transmitter, align the cable with the
notch on the front of the transmitter connector, and push the cable straight down
into the transmitter. To remove the cable, press the latch release on the bottom of
the cable, and pull the cable straight out.
Figure Telemetry-2: SpO
2
Adapter Cable to Transmitter
To enable SpO
2
monitoring in the 90343 digital telemetry multi-parameter
transmitter, choose an averaging interval of 4, 8, or 16, seconds by setting DIP
switches 1 and 2 as explained in Setting SpO2 Data Averaging Period and
Sampling Interval on page -26.
Figure Telemetry-3: Display Zone Full Screen
Figure Telemetry-4: Display Zone Split Screen
Current SpO
2
value (percent) and episodic time of reading. (Time is not
displayed when using continuous mode of operation.)
The bell indicates that alarms are enabled (equal sign turns to a small bell
when SpO
2
alarms are enabled).
SensorWatch bar: shaded area (waveform index WFI) expands up
proportionally to signal strength; horizontal line is minimum signal level.
C
AUTI
O
N:
Never twist the cable.