User's Manual

ECG Troubleshooting Guide
Refer to the Problem Solving sections in this chapter and in the ECG chapter for further monitoring tips.
Clinical Situation Possible Cause Solution
Noisy signal ECG frequency response set to
extended mode
Select monitor mode
Electrodes dry or poor skin adhesion Repeat skin preparation and apply
new, moist electrodes
Baseline wanders Patient moving excessively Secure lead wires by stress-looping to
the patient
Respiration artifact Re-position electrodes
Electrodes dry or poor skin adhesion Repeat skin preparation and apply
new, moist electrodes
Low amplitude ECG Skin improperly prepared Abrade skin and reapply electrodes
Lead selected not providing QRS
complex with greatest amplitude
Select another lead for monitoring
Electrodes could be positioned on
bone or muscle mass
Re-position electrodes
ECG will not learn ECG signal too noisy Check lead wires and electrodes, then
relearn patient rhythm
ECG voltage not within threshold.
ECG VOLTAGE TOO LOW
message may be displayed
Select a different lead or adjust
electrode location
Excessive alarms Electrodes dry or poor skin adhesion Repeat skin preparation and apply
new, moist electrodes
Alarm limits set too close to normal
patient heart rate
Re-adjust alarm limit
Excessive patient movement or
muscle tremor
Reposition electrodes and secure
electrodes with tape if necessary