User's Manual
Table Of Contents
- Directory of Keys - UCW and Ultraview 1700
- Ultraview Digital Telemetry
- General Telemetry Overview
- Cleaning
- Assigning a Telemetry Channel
- Tuning a Receiver for a Bedside
- Entering Patient Information
- Discharging a Patient
- Acknowledging Signal Loss
- Setting Battery Status Alarms
- Controlling Patient-Initiated Recordings
- Telemetry Alarm Message Summary
- ECG Overview
- Setting Up ECG Monitoring
- Display Detail
- Monitoring Paced ECG Patients
- Restoring Default Settings
- Changing the Display Resolution
- Selecting Options for Lead Display
- ECG Alarm Message Summary
- SpO2 Overview (90343 only)
- Setting Up SpO2 Monitoring
- Ensuring Accurate Monitoring
- Setting or Adjusting Alarm Limits
- Setting SpO2 Data Averaging Period and Sampling Interval
- Viewing Pulse Rate
- SpO2 with Intra-Aortic Balloon Pumps
- Using SpO2 with Neonates
- SpO2 Alarm Message Summary
- NIBP Overview (90343 only)
- Setting Up NIBP Monitoring
- Setting Up the ABP Monitor
- Setting or Adjusting Alarm Limits
- Displaying New or Previous Readings
- NIBP Alarm Message Summary
- Alarm Message Summary
- Accessories

Ultraview Digital Telemetry
19
LO RATE ALARM
Displayed during low rate alarms for either 10 seconds or the duration of the
alarm.
NEW DOMINANT
Displayed for 1 minute when a switch to a different dominant ECG morphology
occurs.
NOISY SIGNAL
Displayed in ECG channel 1 when the ECG software suspends processing on
either channel because of excessive noise on the ECG signal. After 10 seconds in
this condition, an alarm tone sounds if ECG alarms are enabled and alarm tones
have not been turned OFF or suspended. This message is displayed for the
duration of the noisy signal condition plus approximately three seconds. The
patient may be moving excessively. Secure the lead wires to the patient.
• Check the electrodes for good skin adhesion.
• Check lead wires at the transmitter for contact.
RUN ALARM
Displayed whenever a RUN of three or more beats is detected and the ABN IN
ROW limit is set lower than or equal to the number of beats in the run. This
message is displayed for either 10 seconds or the duration of the alarm.
V FIB
Displayed whenever ventricular fibrillation is detected. This message is displayed
for either 10 seconds or the duration of the alarm.