User's Manual
Table Of Contents
- Directory of Keys - UCW and Ultraview 1700
- Ultraview Digital Telemetry
- General Telemetry Overview
- Cleaning
- Assigning a Telemetry Channel
- Tuning a Receiver for a Bedside
- Entering Patient Information
- Discharging a Patient
- Acknowledging Signal Loss
- Setting Battery Status Alarms
- Controlling Patient-Initiated Recordings
- Telemetry Alarm Message Summary
- ECG Overview
- Setting Up ECG Monitoring
- Display Detail
- Monitoring Paced ECG Patients
- Restoring Default Settings
- Changing the Display Resolution
- Selecting Options for Lead Display
- ECG Alarm Message Summary
- SpO2 Overview (90343 only)
- Setting Up SpO2 Monitoring
- Ensuring Accurate Monitoring
- Setting or Adjusting Alarm Limits
- Setting SpO2 Data Averaging Period and Sampling Interval
- Viewing Pulse Rate
- SpO2 with Intra-Aortic Balloon Pumps
- Using SpO2 with Neonates
- SpO2 Alarm Message Summary
- NIBP Overview (90343 only)
- Setting Up NIBP Monitoring
- Setting Up the ABP Monitor
- Setting or Adjusting Alarm Limits
- Displaying New or Previous Readings
- NIBP Alarm Message Summary
- Alarm Message Summary
- Accessories

Ultraview Digital Telemetry
17
Table Telemetry-2: Lead Display Options
ECG Alarm Message Summary
Refer to the ECG Problem Solving section in the ECG chapter of this manual for
additional conditions and solutions.
CHECK XX
Displayed in the waveform zone, where XX is the name of the faulted electrode.
The message clears after 60 seconds for V1 – V6 and RL. It is not cleared for limb
lead (RA, LA, LL) faults. If multiple electrodes have faulted, only the highest
priority fault is displayed. The limb leads are highest, followed by RL, and then the
Vx leads.
ABNORMAL/MINUTE ALARM
Displayed whenever the initial abnormal in minute count (A=XX) exceeds the ABN
IN MIN alarm setting. This message is displayed for 10 seconds.
ASYSTOLE
Connected Electrodes
(X)
90343/90347
Valid Lead Vectors
R
LCLL
L
A
R
A
X X X X X V1-6, I, II, III, AVR, AVL,
AVF
XX X X I
XXX X II
XXX X III
X X X X I, II, III, AVR, AVL, AVF
XXXI
XX XII
X X X III
X X X X I, II, III, AVR, AVL, AVF
XXXI
XX X II
X X X III
XXXII
!
• Combinations of leads not included above produce invalid lead
vectors. In general, for at least one valid vector, either RL or C
and two limb leads must be connected.
• The RA lead wire must be connected to the transmitter at all
times. This lead wire also serves as the transmitter’s antenna.
To set or adjust ECG alarms:
1 Touch ECG
2 Touch ALARM LIMITS
3 Touch ECG ALARMS
4 Select ECG ALARM ON
5 Select HI, LO, ABN IN ROW,
and ABN PER MIN
6 Use arrow keys to adjust
7 Touch ST LIMITS CH1 or ST
LIMITS CH2 to adjust ST
segment alarm limits