User's Manual

SPECIFICATIONS
Ultraview
Digital
Telemetry
90343,
90347,
90478-Q,
90478-T,
90478-V,
90479-A
MODULAR RECEIVER
(90478 Options Q, T, and V)
Module Configuration Manager capability (refer
to the Module Configuration Manager chapter
of the Ultraview Care Network Operations
Manual for complete feature specifications)
Trends (with appropriate mainframe option) 24
hours of trended data can be displayed in 1.5-,
3-, 6-, 12-, or 24-hour segments; data is stored
in 1-minute resolution
High Level Analog Output
ECG 1: Used for defibrillator
synchronization
Connector: 3-conductor TT phone jack
Dynamic Range: ±5 mV (±10%), rti
Gain: ECG x 1000 (±5%)
Bandwidth: 0.05 to 30 Hz ±10% (-3 dB)
Module Parameter Count This module counts
as 1 or 2 parameters when computing
parameter capacity for monitors
1 displayed ECG lead = 1 parameter
2 displayed ECG leads = 2 parameters
Receiver Options
The following system configuration options are
available in the 90478
A Basic Arrhythmia: High and low heart rate,
asystole and ventricular fibrillation (2 leads)
B MultiView
I Arrhythmia Enables users to
review trends of abnormals per minute;
provides additional alarms for abnormals per
minute and abnormals in a row (2 leads)
C MultiView II Arrhythmia Enables users to
review the dominant morphology as well as
episodes orclasses of ventricular fibrillation,
ventricular tachycardia (runs), couplets,
single abnormals, tachycardia, pauses,
ventricular and atrio-ventricular pacing;
provides additional alarms for abnormals in a
row, abnormals per minute, and tachycardia
(2 leads)
S ST segment analysis/review/trend (2 leads)
Q Band operation, 608 to 614 MHz
T Band operation, 602 to 608 MHz
V Band operation, 614 to 620 MHz
RECEIVER ELECTRICAL REQUIREMENTS
Power Consumption 5.0 watts
External Indicators LED lights when user
accesses control
RECEIVER PHYSICAL DIMENSIONS
Height: 4.46 in (11.32 cm)
Width: 2.24 in (5.68 cm)
Depth: 7.00 in (17.78 cm)
Weight: 2.4 lbs (1.11 kg)
RECEIVER HOUSING (90479-A)
Accommodates up to 8 modular receivers
HOUSING PHYSICAL DIMENSIONS
Height: 12.0 in (30.5 cm)
Width: 13.5 in (34.3 cm)
Depth: 17.5 in (44.5 cm)
(includes protective cover)
Weight: 32.0 lbs (14.6 kg)
(without modules loaded)
RECEIVER HOUSING POWER
REQUIREMENTS
100120 VAC, 50/60 Hz, 2A; 220240 VAC,
50/60 Hz, 1A
ENVIRONMENTAL REQUIREMENTS
Operating
Temperature: 50
o
to 104
o
F (10
o
to 40
o
C)
Humidity: 10 to 95% (non-condensing)
Altitude: 0 to 10,000 ft (0 to 3,030.3 m)
Storage
Temperature: -40
o
to 149
o
F (-40
o
to 75
o
C)
Humidity: 10 to 100% (non-condensing)
Altitude: -500 to 40,000 ft (-151.5 to
12,121.2 m)
Water Resistance Meets EN60529 IPX1
REGULATORY APPROVALS
All models are ETL listed and meet UL544 or
UL2601-1 standard for electrical safety;
approved by CSA; Models 90343, 90347, and
90478, approved by FCC (47CFR 15.242 and
47CFR) and Industry Canada (RSS - 210,
608-614 MHz operation only)